Dimension RxL Max Integrated Chemistry System

Weight 880 lbs. (400 kg)
Dimensions Width:  62.5 in. (159 cm)
Height:  44 in. (112 cm)
Depth:  32 in. (81 cm)
Operating Temperature 65-85°F (17-30ºC) with a maximum fluctuation of 5ºF (2.8ºC) per hour.  The system requires a maximum of 120 minutes to warm up from a cold start to the incubation temperature.
Relative Humidity 20-80%
Average Noise Output <70 dBA at 1 m while operating
Thermal Output 4778 BTU/hr
Purified Water System is supplied with water system
Electrical Service Outlet

115 vac, 60 Hz, Single Phase, 20A (North America)
230 vac, 50 Hz, Single Phase, 16A (Europe)
230 vac/13A for U.K.
Transient Overvoltage
Installation category II (branch circuit)

 