Tools


Site Explorer

Site Explorer

IgG Subclass 1–4 Immunoassays

The Comprehensive Solution for IgG Determination


Assay characteristics

 N Antiserum IgG/Subclass 1N Antiserum IgG/Subclass 2N Latex IgG3N Latex IgG4
Assay principleLatex-enhanced immunonephelometry
Sample
types
Human serum, EDTA plasma, heparinized plasma
Reference ranges*4.05 – 10.111.69 – 7.860.11 – 0.850.03 – 2.01
Initial measuring range0.85 - 270.35 - 110.033 - 2.10.052 - 3.3
Measuring time6122+6
incl. pre-reaction
2+6
incl. pre-reaction
Once-opened reagent staibilty2 weeks on BN ProSpec® System, 3 days (6 days with evaporation stoppers) on BN™ II System2 weeks on BN ProSpec® System, 3 days (6 days with evaporation stoppers) on BN™ II System2 weeks on BN ProSpec® System, 5 days (10 days with evaporation stoppers) on BN™ II System2 weeks on BN ProSpec® System, 5 days (10 days with evaporation stoppers) on BN™ II System
Calibration frequency4 weeks4 weeks4 weeks4 weeks
Precision
(total CV)
≤ 3% @ 3.6 g/L< 3.0% @ 1.8 g/L≤ 6% @ 0.84 g/L< 3% @ 0.507 g/L

 

*The above reference ranges (2.5th to 97.5th percentile in g/L) were established by testing samples from 279 apparently healthy adults from Central Europe (data on file). Please refer to the download section for pediatric reference ranges.

Did this information help you ?
Thank you for your response
We detected numbers and/or an '@' symbol in your comment. Are you trying to enter a phone number or email address so that we may contact you, please contact us via our email form instead.

Note: to get a direct personal response, please contact us via our email form.