Tools


Site Explorer

Site Explorer

N Latex FLC kappa and N Latex FLC lambda Assays

Add consistency to monoclonal gammopathy testing

IFE, kappa positive N Latex FLC kappa assay Competitive FLC kappa assay
60 60 59 (98.3%)
     
IFE, lambda positive N Latex FLC lambda assay Competitive FLC lambda assay
59 59 (98.3%) 56 (94.9%)
     
IFE, kappa and/or lambda positive N Latex FLC ratio assay Competitive FLC ratio
115 105 (91.3%) 103 (89.6%)

Comparison of N Latex FLC and a competitive FLC assay

(N=152, categorized as abnormal low, normal, abnormal high):

  Agreement rate Cohen’s coefficient
FLC kappa 90.1% 0.76
FLC lambda 81.5% 0.65
FLC ratio 92.2% 0.87

Lot-to-Lot Comparison

 

Comparison of 3 independent reagent lots including 2 different (lot-independent) standard lots (N=95)